手順1に変更
- Meredith Leeによって編集
承認済みの編集 : Meredith Lee
- 前
- 後
- 変更なし
手順ライン
+ | [* black] Put ventilator into diagnostic mode: |
---|---|
+ | [* black] - Begin with ventilator powered off |
+ | [* black] Press ALARM RESET and 100% O2 keys located on the right of the front panel key section (as shown in Figure 8-2) for approximately 5 seconds while turning ventilator power on |
+ | [* black] Touch OK to enter diagnostic mode |