+[title] charging
+[* black] Connect the ventilator back to the AC power.
+[* red] Make sure the battery is properly installed by verifying that the yellow ***Battery (charged) LED*** on the front panel flashes. The flashing LED indicates the battery is being charged.
+[* icon_note] A new battery must be charged for at least 5 hours before being placed into service.
+[* icon_note] The manufacturer recommends that the ventilator’s batteries be fully charged before ventilating a patient. If the batteries are not fully charged and AC power fails, always pay close attention to the level of battery charge.
+[* icon_note] The ventilator charges the battery whenever the ventilator is connected to AC, with or without the ventilator switched on. The ***Battery (charged) LED*** flashes to show that the battery is being charged.