Evita 2 dura Technical Service Bulletin # 20a
Re:
New software version 4.10
Update:
February 13, 2003
Reference Doc:
N/A
Reason:
The following product improvements are available:
- /g183 Quieting of HPSV harmonic resonance in low flow situations on units with the Neoflow option.
- /g183 Elimination of unwanted 'Temp probe ?' alarm errors.
- /g183 Elimination of unwanted Tinsp alarms in mixed modes (SIMV/Psupp) of ventilation in Neoflow.
- /g183 Correction of occasional lack of previous successful device checks being displayed.
- /g183 Correction of intermittent inability to set humidifier configuration.
- /g183 Correction of volume curve disappearance under certain conditions.
- /g183 Elimination of irrelevant display of I:E relation window in NIV.
- /g183 Adjustment of application and cancellation of Testflow during device check.
- /g183 Correction of transmission errors when using Evita Link Option COM2 or COM3 with an external monitor
- /g183 Shortened timeout from 30 to 15 seconds for operator to confirm machine setting changes.
Solution:
Installation of new software version SW 4.10.
Devices affected:
The software update with software version 4.10 is recommended for all Evita 2 dura units.
When:
At next Service Call, Preventive Maintenance, or customer request.
-
Additional Info 1: Any updates/upgrades to machines that were SW version 3.21 or prior will require a clinical in-service or supply of SW 4 Trainer CDROM with P/N 4117256.
-
Additional Info 2: After downloading the new software, perform a device check in stand- by. The message "Loss of data" may occur shortly after first power on after this download and is not a device malfunction. The message will disappear after the internal data backup is completed.
-
Additional Info 3:
See Instructions for TSB #20a- Neoflow units only. (see attachment)
Cost:
The Evita 2 dura SW 4.10 update kit P/N 8414669 including installation is free of charge for Customers in the USA and Canada. Note for US Dräger TSRs: Use Job Type 'Customer Accommodation'.
All other options and their installation are covered by Customer.
Ordering Info: Evita 2 dura software version 4.10 upgrade kit P/N 8414669. The kit consists of Operating Instructions for the Evita 2 dura, Neoflow Operating Instructions, option label, disk with SW 4.10, and new keypad labels.
Distribution:
Dräger Service Personnel and Authorized Service Organizations for CCS products.
If you have any questions, please contact Technical Support by phone at 1-800-543-5047 or by fax at 1-215-721-5789
Dräger Medical, Inc. Technical Product Manager
Instructions for TSB # 20a - (for units fitted with Neoflow TM only)
For units with Neoflow, SW 4.10 allows for optional pulse threshold settings varying between 1 lpm and 5 lpm. 5 lpm is the standard default. It is recommended that the pulse threshold be kept at 5 lpm unless the sound compensation at 5 lpm is inadequate. Other problems such as high PEEP alarms and FiO2 deviations might be more sensitive at lower pulse threshold