Table of contents
| 1 Before use............................................................................... | 3 |
|---|---|
| 2 Device description................................................................. | 7 |
| 3 Preparation/Connection...................................................... | 11 |
| 4 Function check..................................................................... | 13 |
| 5 Preventive maintenance / Storage..................................... | 15 |
| 6 Cleaning................................................................................ | 17 |
| 7 Alarms................................................................................... | 19 |
| 8 Troubleshooting................................................................... | 21 |
| 9 Technical data...................................................................... | 23 |
Table Caption: This table is a table of contents, listing chapter titles and the corresponding page numbers for a manual or document. It outlines the key sections of the document, from initial setup and operation to maintenance and technical specifications.
General information
This symbol on the unit means: Attention, consult accompanying documents.
The device complies with the requirements of the Medical Device Directive 93/42/EEC.
- The Compressor Mini may only be used by authorized personnel. Users must have a thorough knowledge of the unit and follow the instructions in this Operating Manual.
- The owner or user is solely responsible for safety if the equipment is not used as intended, or has been serviced or repaired by personnel not authorized by MAQUET.
- Documentation for the Compressor Mini consists of:
- Operating Manual -
- -Service manual
- -Spare parts list
- -Installation instructions
1. Before use
- Remove the transport safety lock before the compressor is used. Place it in the transport safety lock holder.
- After unpacking the compressor, carry out a function check.
- Pressure data for the Compressor Mini is given in psi or kPa.
- 1 psi = 6,895 kPa
- 1 kPa = approx. 10 cm H 2 O
- 100 kPa = 1 bar ~ 1 atm ~ 1 kgf/cm 2 (kp/cm 2 )
- In addition to the Important information here, always follow local and national regulations.
- The central compressed air supply, when connected to Compressor Mini, must not exceed a pressure of 94 psi (650 kPa/6.5 bar).
1. Before use
Connection
- The function check must always be carried out before the unit is put into use, or once a week when the unit is in continuous use.
- Make sure the mains power and gas fulfill requirements for medical equipment.
- The Compressor Mini must only be connected to a grounded outlet.
Operation
- When in use, the compressor must be placed so that the air vents at the back and on the underside are not covered.
- When in use, the drainage bottle must be checked regularly.
- If the temperature alarm or the pressure alarm is activated during operation, make sure that a fully operational compressor is connected as soon as possible.
Cleaning
The unit should be cleaned at least every