必要な工具と部品

    • Before disassembly, make sure the front wheel breaks are on.

    • Make sure ALL gas supplies are disconnected and the system has been purged, and make sure all electrical supplies have been disconnected.

  1. Ulco Integrus PSV O-Ring Replacement: 手順 4、 1の画像 1
    • With one hand on the ventilator removal handle and one holding the lower half of the ventilator, pull up firmly on the handle so that the mounting slides up the slots in the frame.

  2. Ulco Integrus PSV O-Ring Replacement: 手順 5、 1の画像 1
    • Pull the ventilator housing towards you so that the mounting tabs come out of the frame slots rest the ventilator housing on the plastic skirt at the rear of the workstation as shown in the photo.

  3. Ulco Integrus PSV O-Ring Replacement, Removing the Rear Access Panels in the Ulco Integrus PSV: 手順 6、 1の画像 1
    • Remove the rear plastic cover by turning the single hex key retaining clip to free the cover.

  4. Ulco Integrus PSV O-Ring Replacement: 手順 7、 1の画像 1
    • Remove the four screws mounted to the cover plate using a 4mm Allen key.

  5. Ulco Integrus PSV O-Ring Replacement, O-Ring Replacement: 手順 8、 1の画像 1
    • Pull out the exhaust gas tube (EV910424) and replace the O-ring in the groove on the tube and smear with silicone grease.

  6. Ulco Integrus PSV O-Ring Replacement: 手順 9、 1の画像 1
    • Pull out the drive gas tube (EV9104251) and replace the O-ring in the groove on the tube and smear with silicone grease.

  7. Ulco Integrus PSV O-Ring Replacement: 手順 10、 1の画像 1
    • Pull out the blow off gas tube (EV910426).

  8. Ulco Integrus PSV O-Ring Replacement: 手順 11、 1の画像 1
    • Unscrew the 12 retaining screws on the top plate. Carefully lever the top plate back taking care to ensure that none of the cables connecting it to the body of the ventilator become disconnected.

  9. Ulco Integrus PSV O-Ring Replacement: 手順 12、 1の画像 1
    • Using tweezers, remove the O-ring from inside the blow off gas tube receptacle on the flow block. Smear the O-ring groove with silicone grease and insert the new O-ring.

終わりに

To reassemble your device, follow these instructions in reverse order.

作成者

2人の作成者と共同で作成されました。

Preston Myers

メンバー登録日: 04/09/20

493 ポイント

4のガイドは作成済み

チーム

Cal Poly, Team S7-G13, Paton Spring 2020 Cal Poly, Team S7-G13, Paton Spring 2020人のメンバー

CPSU-PATON-S20S7G13

3 メンバー

8のガイドは作成済み

0 件のコメント

コメントを追加

統計データ:

過去 24時間: 0

過去 7 日: 0

過去 30 日: 1

今までの合計 49