HAMILTON C-1 Bacterial Filter Replacement (Coaxial Circuit Setup)

医療機器
この修理ガイドは医療機器メーカーからの是認を得ておらず、iFixitと医療機器メーカーは提携していません。
  1. HAMILTON C-1 Bacterial Filter Replacement (Coaxial Circuit Setup), Bacterial Filter: 手順 1、 1の画像 1
    • Orient Hamilton-C1 so the right side of the device is facing you.

  2. HAMILTON C-1 Bacterial Filter Replacement (Coaxial Circuit Setup): 手順 2、 1の画像 1
    • Disconnect HME (heat and moisture exchanger) and flex tube

    • Bacterial filter is held within the HME.

    • Use both hands to gently pull and/or twist the flex tube and HME away from one another until fully disconnected.

    • The HME should still be connected to the overall circuit setup.

    • In reassembly, a new HME with a non-contaminated bacterial filter will be reconnected to the flex tube for this step.

  3. HAMILTON C-1 Bacterial Filter Replacement (Coaxial Circuit Setup): 手順 3、 1の画像 1
    • Disconnect HME and circuit setup

    • Use both hands to gently pull and/or twist the flow sensor and HME away from one another until fully disconnected.

    • The HME (with used/contaminated bacterial filter) should now be completely disconnected from all parts.

    • In reassembly, a new HME with a non-contaminated bacterial filter will be reconnected to the flow sensor for this step.

終わりに

To reassemble your device, follow these instructions in reverse order.

作成者

1人の作成者と共同作成されました。

Luisa Chiu

メンバー登録日: 04/06/20

205 ポイント

2のガイドは作成済み

チーム

Cal Poly, Team S7-G11, Paton Spring 2020 Cal Poly, Team S7-G11, Paton Spring 2020人のメンバー

CPSU-PATON-S20S7G11

2 メンバー

6のガイドは作成済み

コメント 1 件

Okay how offen should change out a carbon filter?

Bill Mcleod - 返信

コメントを追加

統計データ:

過去 24時間: 1

過去 7 日: 1

過去 30 日: 30

今までの合計 750