Philips Respironics System One REMstar Blower Unit Replacement

必要な工具と部品

  1. Philips Respironics System One REMstar Blower Unit Replacement, Blower Unit: 手順 1、 1の画像 1
    • Turn the device upside down so that the rubber feet are facing you.

    • Use the TR15 screw head to remove the two 46.5mm length crews located on the back.

    • Gently remove the casing so that the wires are not damaged.

  2. Philips Respironics System One REMstar Blower Unit Replacement: 手順 2、 1の画像 1
    • Using a Torx TR10 screwdriver, remove the seven 13.5 mm screws that are securing the inner casing.

    • Remove the inner casing.

  3. Philips Respironics System One REMstar Blower Unit Replacement: 手順 3、 1の画像 1
    • Remove the motor from the bottom half of the inner casing and set it to the side.

  4. Philips Respironics System One REMstar Blower Unit Replacement: 手順 4、 2の画像 1 Philips Respironics System One REMstar Blower Unit Replacement: 手順 4、 2の画像 2
    • Remove the circular piece by popping it off to separate the circuit board from the top cover.

    • No tools should be necessary to complete this action.

  5. Philips Respironics System One REMstar Blower Unit Replacement: 手順 5、 2の画像 1 Philips Respironics System One REMstar Blower Unit Replacement: 手順 5、 2の画像 2
    • Disconnect the green and red wire connectors from their white connectors to disconnect the blower motor.

終わりに

To reassemble your device, follow these instructions in reverse order.