Philips BiPAP S/T Air Filter Replacement

医療機器
この修理ガイドは医療機器メーカーからの是認を得ておらず、iFixitと医療機器メーカーは提携していません。

ビデオの概要

  1. Philips BiPAP S/T Air Filter Replacement, Air Filter: 手順 1、 1の画像 1
    • Disconnect the device from the power source to stop airflow.

    • To avoid electrical shock, always unplugged the power cord from the wall outlet before cleaning the device.

  2. Philips BiPAP S/T Air Filter Replacement: 手順 2、 2の画像 1 Philips BiPAP S/T Air Filter Replacement: 手順 2、 2の画像 2
    • Remove the filter from the enclosure by gently squeezing the filter in the center and pulling it away from the device.

    • Examine the filter for cleanliness and quality.

    • If the foam filter is torn, replace it. (Only Philips Respironics-supplied filters should be used as replacement filters.)

    • A dirty inlet filter may cause high operating temperatures that may affect device performance. Regularly examine the inlet filter.

  3. Philips BiPAP S/T Air Filter Replacement: 手順 4、 1の画像 1
    • Wash the gray foam filter in warm water with a mild detergent.

    • Rinse thoroughly to remove all detergent residue.

    • Allow the filter to air dry completely before reinstalling it.

    • Never install a wet filter into the device. You must ensure sufficient drying time.

  4. Philips BiPAP S/T Air Filter Replacement: 手順 5、 1の画像 1
    • Reinstall the filter.

    • To install, insert gray foam filter into the filter area.

    • A properly installed, undamaged filter is required for proper operation.

終わりに

To reassemble your device, follow these instructions in reverse order.

作成者

1人の作成者と共同作成されました。

Sophia James

メンバー登録日: 04/07/20

193 ポイント

1ガイドが承認済み

チーム

Cal Poly, Team S7-G3, Paton Spring 2020 Cal Poly, Team S7-G3, Paton Spring 2020人のメンバー

CPSU-PATON-S20S7G3

3 メンバー

3のガイドは作成済み

2 件のコメント

In India corona virus control in a earlier situation and now its spread in many countries again '''bipap machines''' are necessary for many peoples in this situation but you have to be careful from your side as well btw thanks for informative article.

mers - 返信

Why is there a gray blue material coming out on the output filter. It has been doing for over two years. Is it killing me. My bi pa was recalled over a year ago, when are they going to replace it ?

EDward Stone - 返信

コメントを追加

統計データ:

過去 24時間: 1

過去 7 日: 1

過去 30 日: 16

今までの合計 1,303