How to Test the Valve System on a Fisher Paykel Neopuff

医療機器
この修理ガイドは医療機器メーカーからの是認を得ておらず、iFixitと医療機器メーカーは提携していません。
    • Set the gas supply to 5 L/min.

  1. How to Test the Valve System on a Fisher Paykel Neopuff: 手順 2、 1の画像 1
    • Completely close the Peak Inspiratory Pressure control and the Max Pressure control by turning both knobs completely clockwise.

  2. How to Test the Valve System on a Fisher Paykel Neopuff: 手順 3、 1の画像 1
    • Close the bleed valve and check that the gauge reads at least 60 cmh2O.

    • Set the gas supply to 15 L/min.

  3. How to Test the Valve System on a Fisher Paykel Neopuff: 手順 5、 1の画像 1
    • Close the bleed valve and check that the gauge reads no higher than 80 cmH2O.

    • Set the gas supply to 10 L/min.

    • Close the bleed valve and turn the Max Pressure Relief valve until the manometer reads 40 cmh2O.

    • Check that the manometer needle rises and falls smoothly.

    • Gently rotate the Max Pressure Relief cover over the Max Pressure Relief knob.

  4. How to Test the Valve System on a Fisher Paykel Neopuff: 手順 10、 1の画像 1
    • Reset the peak inspiratory pressure to 20 cmh2O and turn off the gas flow. Testing is now complete. If the F&P Neopuff/Perivent Infant Resuscitator fails any of these tests, the valve assembly should be regarded as faulty and replaced with a new valve assembly (see the parts list (section 4.1.2) for part numbers).

終わりに

To reassemble your device, follow these instructions in reverse order.

ある他の人がこのガイドを完成しました。

作成者

2人の作成者と共同で作成されました。

Constantine Yap

メンバー登録日: 05/04/20

364 ポイント

2のガイドは作成済み

チーム

Cal Poly, Team S1-G6, Zimmerman Spring 2020 Cal Poly, Team S1-G6, Zimmerman Spring 2020人のメンバー

CPSU-ZIMMERMAN-S20S1G6

4 メンバー

27のガイドは作成済み

0 件のコメント

コメントを追加

統計データ:

過去 24時間: 0

過去 7 日: 0

過去 30 日: 4

今までの合計 497