Cleaning the Ulco Integrus PSV

医療機器
この修理ガイドは医療機器メーカーからの是認を得ておらず、iFixitと医療機器メーカーは提携していません。
  1. Cleaning the Ulco Integrus PSV, Cleaning the Ulco Integrus PSV: 手順 1、 2の画像 1 Cleaning the Ulco Integrus PSV, Cleaning the Ulco Integrus PSV: 手順 1、 2の画像 2
    • Absorber

    • Loosen the knurled screws and remove the clear lid.

  2. Cleaning the Ulco Integrus PSV: 手順 2、 1の画像 1
    • Remove the silicone shutters by gently lifting from the base, NOT from the flaps.

  3. Cleaning the Ulco Integrus PSV: 手順 3、 1の画像 1
    • To release and remove the canister(s), undo the handle latch by turning sideways.

  4. Cleaning the Ulco Integrus PSV: 手順 4、 1の画像 1
    • Push the release handle down to vertical until canisters are free.

  5. Cleaning the Ulco Integrus PSV: 手順 5、 1の画像 1
    • Push canister(s) down then out of the cradle. Then dispose of the used soda lime (see suppliers recommended procedure for disposal of soda lime).

    • Wash all the absorber components. Do NOT use caustic cleaning fluids.

    • The Canister(s) and silicone shutters can be autoclaved.

    • The Absorber head can be put through a washer at 80 deg.C

    • Dry thoroughly before assembly, low pressure warm air should be passed through the head by attaching a hose to the expiratory port of the absorber.

  6. Cleaning the Ulco Integrus PSV: 手順 6、 2の画像 1 Cleaning the Ulco Integrus PSV: 手順 6、 2の画像 2
    • Bellows Assembly

    • Loosen the four knurled screws (1) and remove the bellows canister.

  7. Cleaning the Ulco Integrus PSV: 手順 7、 1の画像 1
    • Remove the bellows assembly by rotating counter-clockwise. The bellows canister may be washed or autoclaved.

  8. Cleaning the Ulco Integrus PSV: 手順 8、 1の画像 1
    • The base disk (mushroom) should be removed from the bellows assembly before washing. Pull the latex-free rubber bag from the delrin base disk. The bellows may be washed or autoclaved. The base component (mushroom) should not be autoclaved but can be put through a washer at 80°C.

    • Dry all components thoroughly before re-assembly.

    • Low pressure warm air should be passed through the absorber head by attaching a hose to the scavenge port

終わりに

To reassemble your device, follow these instructions in reverse order.

ある他の人がこのガイドを完成しました。

作成者

2人の作成者と共同で作成されました。

Jacob Henriksen

メンバー登録日: 04/15/19

232 ポイント

1ガイドが承認済み

チーム

Cal Poly, Team S7-G13, Paton Spring 2020 Cal Poly, Team S7-G13, Paton Spring 2020人のメンバー

CPSU-PATON-S20S7G13

3 メンバー

8のガイドは作成済み

0 件のコメント

コメントを追加

統計データ:

過去 24時間: 0

過去 7 日: 0

過去 30 日: 1

今までの合計 77